Siemens SITRANS LR460 HART Operating Instructions Manual page 109

Radar transmitters
Hide thumbs Also See for SITRANS LR460 HART:
Table of Contents

Advertisement

LCD
Meaning (cont' d )
display
Input parameters High
Calibration Point and Low
S:10
Calibration Point are the
same.
Internal temperature sensor
S:11
failure.
Internal temperature of the
device has exceeded
specifications: it is operating
S:12
*
outside its temperature
range.
Upper and lower input values
(Process Value Scale) for
S:14
*
AIFB1 are the same.
Upper and lower input values
(Process Value Scale) for
S:15
AIFB2 are the same.
Calibration interval as
defined in Maintenance
S: 17
Required Limit has expired.
Calibration interval as
defined in Maintenance
S: 18
Demanded Limit has expired.
DSP could not be initialized
S: 19
The TVT Shaper could not be
S: 20
recovered.
The AFES profile could not be
S: 21
recovered.
A5E42035355
Corrective Action
SITRANS LR460 – INSTRUCTION MANUAL
Check calibration settings of
device. Ensure settings for
High Calibration Point and
Low Calibration Point are
different.
Return the device to the
factory.
Lower the ambient
temperature enough to cool
the device.
Fault code will persist until a
manual reset is performed
using PDM or the LCD
interface.
Check configuration for
AIFB1. Ensure that Upper
Value and Lower Value
(Process Value Scale) are not
the same.
Check configuration for
AIFB2. Ensure that Upper
Value and Lower Value
(Process Value Scale) are not
the same.
Perform calibration.
Perform calibration.
If fault persists, return device
to manufacturer
If fault persists, return device
to manufacturer.
If fault persists, return device
to manufacturer.
Byte
Bit
(cont' d )
2
3
4
1
6
7
1
2
2
3
4
5
Page B-103