Siemens SITRANS LR260 Operating Instructions Manual

Radar, hart
Hide thumbs Also See for SITRANS LR260:
Table of Contents

Advertisement

Radar Transmitters
SITRANS LR260 (HART)
Operating Instructions 12/2012
SITRANS

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SITRANS LR260

  • Page 1 Radar Transmitters SITRANS LR260 (HART) Operating Instructions 12/2012 SITRANS...
  • Page 2 The user is responsible for all changes and repairs made to the device by the user or the user’s agent.  All new components are to be provided by Siemens Milltronics Process Instruments.  Restrict repair to faulty components only.
  • Page 3: Table Of Contents

    Safety Notes ............................1 Safety marking symbols ......................1 FCC Conformity ............................1 CE Electromagnetic Compatibility (EMC) Conformity ..............2 The Manual ............................3 Technical Support ..........................3 SITRANS LR260 Overview ....................5 Specifications ........................6 Power............................. 6 Performance..........................6 Interface ............................7 Mechanical (Easy Aimer model) ..................... 7 Mechanical (Threaded Connection model) ................
  • Page 4 Requesting an Echo Profile ......................34 Device Address ...........................34 Auto False Echo Suppression ......................34 Level application example ........................35 Operating via SIMATIC PDM ..................36 Functions in SIMATIC PDM ......................36 Electronic Device Description (EDD) ..................37 Configuring a new device .......................37 Quick Start Wizard via SIMATIC PDM ..................37 Changing parameter settings using SIMATIC PDM ..............41 Parameters accessed via pull-down menus ..............41 Parameter Reference .....................49...
  • Page 5 Process Pressure/Temperature derating curves ..............102 Horn Antenna .............................103 Loop power ............................104 Typical Connection Drawing ....................104 Allowable operating area of SITRANS LR260 ..............104 Startup Behavior ........................105 Appendix E: Application Example ................107 Flour in steel storage vessel, level measurement ..............107 Appendix F: HART Communications .................109 SIMATIC PDM ...........................109...
  • Page 6 Notes...
  • Page 7: Safety Notes

    FCC Conformity US Installations only: Federal Communications Commission (FCC) rules WARNING: Changes or modifications not expressly approved by Siemens Milltronics could void the user’s authority to operate the equipment. Notes: • This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
  • Page 8: Ce Electromagnetic Compatibility (Emc) Conformity

    Testing and measurement techniques – Immunity to EN61000-4-6:2004 conducted disturbances, induced by radio-frequency fields. Electromagnetic Compatibility (EMC) Part 4-8: EN61000-4-8:2001 Testing and measurement techniques – Power frequency magnetic field immunity test. Page 2 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 9: The Manual

    SITRANS LR260. • This manual applies to the SITRANS LR260 (HART) only. This manual will help you set up your SITRANS LR260 for optimum performance. We always welcome suggestions and comments about manual content, design, and accessibility.
  • Page 10 F microFarad s microsecond Pressure Equipment Directive safety approval pico Farads parts per million Primary Variable measured value Secondary Variable equivalent value sensitivity Time Varying Threshold threshold Input voltage Output voltage Page 4 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 11: Sitrans Lr260 Overview

    1,000,000 applications worldwide (ultrasonic and radar). Programming SITRANS LR260 is very easy to install and configure via a graphical local user interface (LUI). You can modify the built-in parameters either locally via the Siemens infrared handheld programmer, or from a remote location via SIMATIC PDM.
  • Page 12: Specifications

    Specifications Siemens Milltronics makes every attempt to ensure the accuracy of these Note: specifications but reserves the right to change them at any time. Power Nominal 24 V DC with Loop For other configurations, see the chart under max. 550 Ohm loop resis-...
  • Page 13: Interface

    Display quality will be degraded in temperatures below –25 °C (–13 °F) and above +65 °C (+149 °F). Universal flange mates with EN 1092-1 (PN16)/ASME B16.5 (150 lb)/JIS 2220 (10K) bolt hole pattern. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 7...
  • Page 14: Mechanical (Threaded Connection Model)

    Process Pressure/Temperature derating curves more detail, or for other configurations, see Process Pressure/Temperature derating curves page 102, and on page 102. Pressure rated version for maximum process temperature of 80 °C (176 °F) Page 8 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 15: Programmer (Infrared Keypad)

    FM/CSA Class I, Div 2, Groups A, B, C, D T5 SABS ARP0108 Ex ia IIC T4 Ga Programmer (infrared keypad) Siemens Milltronics Infrared IS (Intrinsically Safe) Hand Programmer for hazardous and all other locations (battery is non-replaceable) • approval ATEX II 1 G, EEx ia IIC T4, cert.
  • Page 16: Dimensions

    2" horn: 163 mm (6.4") 3" horn: 228 mm (9.0") horn 4" horn: 285 mm (11.2") 49 mm 2" horn (1.9") 74.5 mm 3" horn (2.9") 97.5 mm 4" horn (3.8") Page 10 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 17: Threaded Horn Antenna With Extension

    28 mm (1.1") 100 mm (4.0 ") ref. horn extension point process connection horn horn O.D. Threaded Horn dimensions See table under on page 12 for horn height to sensor reference point. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 11...
  • Page 18: Threaded Horn Dimensions

    Height from bottom of horn to sensor reference point as shown: see Horn Antenna with extension on page 11. Polarization reference point – 3dB in the direction of the polarization axis (see on page 15 for an illustration). Page 12 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 19: Installation

    • Installation shall only be performed by qualified personnel and in accordance with local governing regulations. • SITRANS LR260 is to be used only in the manner outlined in this manual, otherwise protection provided by the device may be impaired.
  • Page 20: Mounting Location

    3" horn = 10° Easy Aimer possible. (See 4" horn = 8° on page 16.) Extensions are not recommended for applications where there may be excessive visible vibration. Please consult factory for more information. Page 14 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 21: Orientation In A Vessel With Obstructions

    Orientation in a vessel with obstructions Polarization reference point For best results on a vessel with polarization axis obstructions, orient the front or back of the device toward the obstructions. polarization reference point display 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 15...
  • Page 22: Easy Aimer

    Easy Aimer loosen the Easy Aimer ball locking bolts and gently reposition the enclosure. Direct SITRANS LR260 so the horn antenna is pointed at an angle perpendicular to the material surface, if possible. (As a guide, aim the beam at a point approximately 2/3 of the way across the tank diameter.)
  • Page 23: Air Purging System (Optional)

    • Some dust particles are highly abrasive and can be drawn into the inside of the horn during purge cleaning, damaging the internal PTFE emitter of the antenna. A replacement kit is available from your local Siemens Milltronics representative. • It is the customer’s responsibility to ensure that any vacuum or pressure in the measured vessel is maintained, considering the hole that passes through the process connection and SITRANS LR260 antenna system.
  • Page 24: Universal Slotted Flange

    0.38" 9.53" 9.45" 0.45" 150 mm (285 mm) (9.65 mm) (242 mm) (240 mm) (11.5 mm) Universal flange mates with EN 1092-1 (PN16)/ASME B16.5 (150 lb)/JIS 2220 (10K) bolt hole pattern. Page 18 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 25: Dust Cap (Optional)

    Hand tighten the adjustable clamp supplied to secure the cap. Use a screwdriver or nut driver to tighten the clamp screw until the clamp provides an air-tight seal. It is critical to ensure no moisture can be trapped inside. Note: 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 19...
  • Page 26: Wiring

    Unscrew the cover for access to the terminals. Use a screwdriver for added leverage, if necessary. plug cable gland (or NPT cable entry) Depending on the approval rating, glands and plugs may be supplied with your instrument. Page 20 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 27 Depending on the system design, the power supply may be separate from the PLC, or integral to it. Loop resistance (total of cable resistance plus 250 Ohm [resistor]) must be less than 550 Ohm for the device to function properly. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 21...
  • Page 28: Intrinsically Safe Wiring

    Device nameplate (ATEX/IECEX/INMETRO/C-TICK) The ATEX certificate listed on the nameplate can be downloaded from the product page of our website at: www.siemens.com/LR260. Go to Support > Approvals / Certificates. The IECEx certificate listed on the nameplate can be viewed on the IECEx website. Go to: http://iecex.iec.ch and click on Ex Equipment Certificates of Conformity then enter the...
  • Page 29 9001/01-280-100-10 (Single Channel) Stahl 9002/01-280-110-10 (Dual Channel) How to select a passive barrier for SITRANS LR260 To make sure that the barrier safety description is suitable for the LR260 Intrinsically Safe (IS) input parameters, carry out the following calculations: Re-e max.
  • Page 30: Non-Incendive Wiring (Us/Canada Only)

    2. Non-incendive wiring (US/Canada only) FM/CSA Class 1, Div 2 connection drawing number A5E03745541 can be downloaded from the product page of our website at: www.siemens.com/LR260. Go to Support > Installation Drawings > Level Measurement > Continuous - Radar. Startup Behavior...
  • Page 31 For applications that use the purge facility, the user shall provide a means to ensure that combustible dust from the hazardous area cannot enter the purge supply in such a way as to compromise the area classification. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 25...
  • Page 32: Quick Start Via Local Operation

    Quick Start via local operation SITRANS LR260 carries out its level measurement tasks according to settings made via parameters. The settings can be modified locally via the Local User Interface (LUI) which consists of an LCD display and a handheld programmer.
  • Page 33 • A deeper band indicates fewer items. Parameter view parameter name parameter number parameter value/ selection Edit view 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 27...
  • Page 34: Handheld Programmer (Part No. 7Ml1930-2Aj)

    RIGHT arrow opens PROGRAM Opens the top level menu. mode. UP or DOWN arrow toggles between LCD displays measured value in either linear units or linear units and percent. percent. Page 28 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML1998KE03...
  • Page 35: Programming Sitrans Lr260

    • Do not use the handheld programmer at the same time as SIMATIC PDM or erratic operation may result. • SITRANS LR260 automatically returns to Measurement mode after a period of inactivity in PROGRAM mode (between 15 seconds and 10 minutes, depending on the menu level).
  • Page 36 Edit mode. The current selection is highlighted. Scroll to a new selection. Press RIGHT arrow to accept it The LCD returns to parameter view and displays the new selection. Page 30 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML1998KE03...
  • Page 37 Erases the display. editing Decimal Numeric Enters a decimal point. point editing Plus or Numeric Changes the sign of the entered value. minus sign editing Numeric Numeral Enters the corresponding character. editing 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 31...
  • Page 38: Quick Start Wizard Via The Handheld Programmer

    Use a setting just faster than the maximum filling or emptying rate (whichever is greater). 1.3. Units Select the units for the Quick Start variables (high and low calibration point, and level, distance, or space) M, CM, MM, FT, IN Options Default: M Page 32 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML1998KE03...
  • Page 39 YES, NO, DONE (Display shows DONE when Quick Start is successfully Options completed) Press Mode to return to Measurement mode. SITRANS LR260 is now ready to operate. The point from which High and Low Calibration points are referenced: see Dimensions: on page 10.
  • Page 40: Requesting An Echo Profile

    80. Auto False Echo Suppression If SITRANS LR260 displays a false high level, or the reading is fluctuating between the correct level and a false high level, you can use the Auto False Echo Suppression 2.2.5. TVT (Auto False Echo Suppression) parameters to prevent false echo detection.
  • Page 41: Level Application Example

    Fill rate = 0.09 m/minute (Low Cal Pt. minus High Cal Pt.) / fastest fill/empty time = (16 m – 0.5 m) / 180 min. = 15.5 m /180 min. = 0.09 m/min. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 35...
  • Page 42: Operating Via Simatic Pdm

    Note: starting on page 49. SIMATIC PDM is a software package used to commission and maintain SITRANS LR260 and other process devices. Please consult the operating instructions or online help for details on using SIMATIC PDM. (You can find more information at www.fielddevices.com: go to Products and Solutions >...
  • Page 43: Electronic Device Description (Edd)

    Electronic Device Description (EDD) SITRANS LR260 requires the EDD for SIMATIC PDM version 6.0 with SP3 or Note: higher. You can locate the EDD in Device Catalog, under Sensors/Level/Echo/Siemens Milltronics/ SITRANS LR260. Check the product page of our website at: www.siemens.com/LR260, under Downloads, to make sure you have the latest version of SIMATIC PDM, the most recent Service Pack (SP) and the most recent hot fix (HF).
  • Page 44 Note: The layout of the dialog boxes shown may vary according to the resolution setting for your computer monitor. Click on NEXT to accept the default values. (Description, Message, and Installation Date fields can be left blank.) Page 38 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 45 Step 2 – Application Type Select the application type and the operation, then click on NEXT. Step 3 – Range Setup Set the parameters, and click on NEXT. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 39...
  • Page 46 Check parameter settings, and click on BACK to return and revise values, or TRANSFER to transfer values to the device. The message Quick Start was successful will appear. Click on OK, then click on OK again to synchronize with the device. Page 40 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 47: Changing Parameter Settings Using Simatic Pdm

    • Clicking on Cancel during an upload from device to SIMATIC PDM will result in some parameters being updated. Launch SIMATIC PDM, connect to SITRANS LR260, and upload data from the device. Adjust parameter values in the parameter value field then press Enter. The status fields read Changed.
  • Page 48 • You can save a profile or delete a saved profile. • After saving a profile, open menu View – Echo profile. save delete Page 42 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 49 • Double click on each axis and record the Xscale and Data Scale values, so that you can restore the default view by resetting to these values. Open the menu View – Trend trend line 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 43...
  • Page 50 • Before adjusting these parameters, rotate the instrument for best signal (lower false-echo amplitude). SITRANS LR260 first learns the echo profile. Then the learned profile, or part of it, is used to Auto False Echo Suppression screen out false echoes. (See on page 98 for a more detailed explanation.)
  • Page 51 Change the position of the cursor on the TVT curve using the Point+ and Point– buttons: raise and lower the curve using Offset+ and Offset–. • Alternatively, enter values for Point and Offset directly into the dialog boxes. • Press Transfer to Device. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 45...
  • Page 52 Click on Read, to see the effects of your modification. If desired, activate alerts for either or both of Maintenance Required/Maintenance Demanded. Click on Snooze to add a year to the Total Expected Device Life. Page 46 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 53 Open the menu Device – Select AnalogOut. The Select AnalogOut window displays the current setting: click on OK. Select the desired option, and click on OK. The Select AnalogOut window displays the new setting: click on OK. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 47...
  • Page 54: Hart Communication

    Open the menu Device – HART Communication. How to Set Device Address The default address for the device is 0. To reset the address via the handheld programmer see 5. 1 .Device Address on page 80. Page 48 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 55: Parameter Reference

    TVT shaper Maintenance 75 to 77 Toolbar Device Reset Status bar Configuration Flag Reset Select AnalogOut D/A Trim Simulate AO Update HART Communication Visible only after an echo profile has been saved. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 49...
  • Page 56: Quick Start Wizard

    Calibration Point (process empty level). Distance to material surface referenced from High Options SPACE Calibration Point (process full level). Distance to material surface referenced from Sensor DISTANCE Reference Point. Default: DISTANCE Page 50 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 57 In order to save the Quick Start settings it is necessary to select Yes and apply changes. YES, NO, DONE Options Display shows DONE when Quick Start is successfully completed. • Press Mode to return to Measurement mode. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 51...
  • Page 58: Setup

    Corresponds to the software or firmware that is embedded in the SITRANS LR260. Read only. 2.1.2. Loader Revision (LOADER REV.) Corresponds to the software used to update the SITRANS LR260. Read only. 2.1.3. Hardware Revision (HARDWARE REV.) Corresponds to the electronics hardware of the SITRANS LR260. Read only.
  • Page 59 When setting the High Calibration Point value, note that echoes are ignored within Near Range (2.2.1. 1 1). Dimensions: The point from which level measurement is referenced. See on page 10. The distance from sensor reference point to the target. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 53...
  • Page 60 Units are defined in Sensor Units. Range: -99 999 to 99 999 Values Default: 0 m The value produced by the echo processing which represents the distance from teh sensor reference point to the target. Page 54 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 61 Range: 5 to 95% Values Default: 70% Related 2.2.4.1.2. Echo Position Parameters When SITRANS LR260 is using Rising Edge, higher values will cause lower level readings and lower values will cause higher level readings. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 55...
  • Page 62 2.2.4.2.5. Down Sampling Specifies the number of consecutive echoes that must appear below the echo currently selected, before the measurement is accepted as valid. Range: 1 to 50 Values Default: 5 Page 56 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 63 For example, to filter out a spike with 500 mm width, enter 6 or 7 (the closest integer product of 500 x 0.013). When a value is keyed in, the nearest acceptable value is entered. Echo Lock on page 97 for more detail. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 57...
  • Page 64 Displays the average ambient noise (in dB above 1 μV rms) of a noise profile (view only). Noise level is a combination of transient noise and receiving circuitry. After a measurement, the values from the previous noise shot will be displayed. Page 58 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 65 When SITRANS LR260 is located in the center of the vessel, the TVT hover level may be lowered to increase the confidence level of the largest echo.
  • Page 66 Press RIGHT ARROW to open Edit mode. Enter the new value and press RIGHT ARROW to accept it. 2.2.5.2. Auto False Echo Suppression 2.2.5.4. Shaper Mode Adjusts the TVT curve at a specified range. Options Page 60 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 67 Range: –50 to 50 dB Values Default: 0 dB 2.2.6.4. Shaper 28-36 Range: –50 to 50 dB Values Default: 0 dB 2.2.6.5. Shaper 37-40 Range: –50 to 50 dB Values Default: 0 dB 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 61...
  • Page 68 2.2.7.2. Fill Rate (FILL RATE/MIN) Defines the maximum rate at which the reported sensor value is allowed to increase. Allows you to adjust the SITRANS LR260 response to increases in the actual material level. Fill Rate is automatically updated whenever Response Rate is altered.
  • Page 69: Output

    Range: 3.6 mA to 22.6 mA 2.3.1.2. mA Output Function (mA OUTPUT FUNC.) Alters the mA output/measurement function and allows the output to be set independently. Manual Level Options Space * Distance 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 63...
  • Page 70 (Operation resets mA Output Function to the same value.) • The mA Output Function controls the primary value and the loop current for the SITRANS LR260. Use caution when changing the mA Output function when connected to a HART network.
  • Page 71 4 mA setpoint can be set less than 0 m, and 20 mA setpoint can be set greater than 30 m when using handheld programmer. Recommended range is 0 to 30 m. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 65...
  • Page 72: Fail-Safe

    Defines a user-defined level to report when the Fail-safe timer expires. Range: 3.6 mA to 22.6 mA Values Default: 22.6 mA This value can be used only if Fail-safe Material Level is set to Note: VALUE. Page 66 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 73: Diagnostics

    3.14.2. Maximum Internal Temperature (MAX. INTERN. TEMP.) Read Only. Displays (in degrees C) the maximum temperature recorded by the internal electronics. The high and low values are maintained over a power cycle. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 67...
  • Page 74: Remaining Device Lifetime (Remain. Dev. Life.)

    Click on Snooze to add a year to the Total Expected Device Life. In device, Maintenance parameters are displayed in years. In SIMATIC PDM, mainte- nance parameter units are user selectable. Page 68 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 75 Warning Limits 1 and 2 (Maintenance Required and Options Maintenance Demanded Limits) * OFF a) First set the limit values in Maintenance Required Limit (3.15.3.)/ Maintenance Demanded Limit (3.15.4.). b) Select the desired Maintenance Alert Activation option. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 69...
  • Page 76 In the Device Lifetime section, click on Acknowledge Warnings. To acknowledge an alert via the handheld programmer: a) Press RIGHT arrow twice to open parameter view and activate Edit Mode. b) Press RIGHT arrow to acknowledge the alert. Page 70 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 77: Remaining Sensor Lifetime (Remain Sens. Life.)

    • Via the handheld programmer, manually reset Total Sensor Operating Time to zero. In device, Maintenance parameters are displayed in years. In SIMATIC PDM, mainte- nance parameter units are user selectable. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 71...
  • Page 78 Warning Limits 1 and 2 (Maintenance Required and Options Maintenance Demanded Limits) * OFF a) First set the limit values in Maintenance Required Limit/Maintenance Demanded Limit. b) Select the desired Maintenance Alert Activation option. Page 72 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 79: Service

    To perform a reset to factory defaults via SIMATIC PDM, open the menu Device – Device Reset and click on Factory Defaults. 4.2. Manufacture Date (MANUF. DATE) The date of manufacture of the SITRANS LR260 (yy mm dd). 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL...
  • Page 80: Lcd Fast Mode

    Read only. The number of power cycles that have occurred since manufacture. Via SIMATIC PDM, open the menu View – Wear. 4.11. Memory Test (MEM. TEST) Allows verification of the RAM, EEPROM, and Flash memory of the SITRANS LR260. IDLE No test in progress.
  • Page 81: Service Interval

    • Via the handheld programmer, manually reset Time Last Serviced to zero. In device, Maintenance parameters are displayed in years. In SIMATIC PDM, mainte- nance parameter units are user selectable. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 75...
  • Page 82 Service Interval parameters in SIMATIC PDM. User-configurable recommended time between product inspections Units : hours, days, years Range: 0 to 20 years Values Default: 1 year Units are selectable only via SIMATIC PDM. Page 76 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 83: Calibration Interval (Calib. Interval)

    After modifying values/units as required, click on Write to accept the change, and Read to view the effect of the change. In device, Maintenance parameters are displayed in years. In SIMATIC PDM, mainte- nance parameter units are user selectable. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 77...
  • Page 84 If Remaining Lifetime is equal to or less than this limit, the device sets Maintenance Demanded. Range: 0 to 20 years Values Default: 0.019 years a) Modify limit values as required. b) Set Maintenance Alert Activation to the desired option. Page 78 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 85 In the Calibration Schedule Status section, click on Acknowledge Warnings. To acknowledge an alert via the handheld programmer: a) Press RIGHT arrow twice to open parameter view and activate Edit Mode. a) Press RIGHT arrow to acknowledge the alert. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 79...
  • Page 86: Communication

    Edit mode. Key in a new value and press RIGHT arrow to accept it. 5.2. Communication Control SITRANS LR260 can only reset this parameter via the handheld Note: programmer. Enables /disables the read/write access to parameters via remote communications.
  • Page 87: Security

    • After a new value is stored in Unlock Value (6.2), that value will be recalled after a reset to Factory Defaults. • Consult your Siemens Milltronics representative if you have forgotten the unlock value. Stores the value to be entered in 6.1. Lock to unlock programming. If Lock is on,...
  • Page 88 Page 82 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 89: Appendix A: Alphabetical Parameter List

    Fail-safe Level (2.4.4.) Fail-safe Material Level (FAILSAFE MAT. LE) (2.4.2.) Fail-safe Timer (2.4.1.) Far Range (2.2.1.13.) Fill Rate (FILL RATE/MIN) (2.2.7.2.) Firmware Revision (FIRMWARE REV.) (2.1.1.) Hardware Revision (HARDWARE REV.) (2.1.3.) 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 83...
  • Page 90 Reform Echo (2.2.4.3.7.) Response Rate (2.2.7.1.) Sensor Offset (2.2.1.25.) Sensor Units (2.2.1.4.) Shaper Mode (2.2.5.4.) Space Measurement (SPACE MEAS.) (2.2.8.3.) For diagnostic purposes TVT Hover Level (2.2.5. 1 .) TVT Type (2.2.5.6.) Page 84 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 91 Parameter Name (Parameter Number) Page Number Unlock value (6.2.) Up Sampling (2.2.4.2.4.) Window (2.2.4.2.6.) 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 85...
  • Page 92 Page 86 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 93: Appendix B: Troubleshooting

    FAQs for SITRANS LR260, or contact your local Siemens Milltronics representative. Specifically: If you try to set a SITRANS LR260 parameter via remote communications, but the parameter remains unchanged: • Some parameters can only be changed when the device is not scanning. Try putting the device in PROGRAM mode using the operating mode function.
  • Page 94: Device Status Icons

    • Simulation or substitute value • Communication is good; device is in simulation mode or works with substitute values. • Out of operation • Communication is good; device is out of action. Page 88 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 95: General Fault Codes

    Siemens repre- sentative. S: 2 Unable to collect profile because of Repair required: contact your local a power condition that is outside of Siemens representative. the operating range of the device. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 89...
  • Page 96 Inspect for heat-related damage and contact your local Sie- mens representative if repair is required. Fault code will persist until a manual reset is performed using PDM or the LCD interface. Page 90 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 97 Factory calibration for the device Repair required: contact your local has been lost. Siemens representative. S: 35 Factory calibration for the device Repair required: contact your local has been lost. Siemens representative. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 91...
  • Page 98 EEPROM corrupt. Repair required: contact your local Siemens representative. S: 50 EEPROM corrupt. Repair required: contact your local Siemens representative. S: 51 EEPROM corrupt. Repair required: contact your local Siemens representative. Page 92 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 99: Operation Troubleshooting

    3.14. 1 . Current Internal antenna malfunction: • check Display shows Temperature (CURR. INTERN. • temperature too high S: 0 LOE TEMP.) • physical damage • relocate 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 93...
  • Page 100 (minimum value depends on than material level • multiple echoes antenna type) processed • raise SITRANS LR260 2.2.4. 1 . 1 . Algorithm • ensure is set to F (First echo) Page 94 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 101: Appendix C: Maintenance

    Appendix C: Maintenance SITRANS LR260 requires no maintenance or cleaning under normal operating conditions. Under severe operating conditions, the horn antenna may require periodic cleaning. If cleaning becomes necessary: • Note the antenna material and the process medium, and select a cleaning solution that will not react adversely with either.
  • Page 102: Appendix D: Technical Reference

    Appendix D: Technical Reference Principles of Operation SITRANS LR260 is a 2-wire 25 GHz pulse radar level transmitter for continuous monitoring of solids . Radar level measurement uses the time of flight principle to determine distance to a material surface. The device transmits a signal and waits for the return echo.
  • Page 103: Time Varying Threshold (Tvt) Curves

    • Lock Off: SITRANS LR260 responds immediately to a new selected echo (within the restrictions set by the Maximum Fill / Empty Rate), but measurement reliability is affected. Echo Position Detection 2.2.4.1.2.
  • Page 104: Auto False Echo Suppression

    Measurement Range Near Range 2.2.1.12. Near Range programs SITRANS LR260 to ignore the zone in front of the antenna. The default blanking distance is 50 mm (1.97") from end of horn antenna. 2.2.5.2. Near Range allows you to increase the blanking value from its factory default. But...
  • Page 105: Far Range

    20 mA 4 mA When SITRANS LR260 is put into PROGRAM mode it stops responding to the process. It stores the most recent measurement, and holds the associated readings and mA signal output. The instrument reverts to the parameter last addressed during the previous program session.
  • Page 106: Damping

    The reading and mA output default to the last measurement taken. The reading and associated outputs migrate to the current process level at a rate controlled by the response rate (2.2.7.1). If SITRANS LR260 is left in PROGRAM mode for 10 minutes without input, it automatically reverts to Measurement mode. Damping A damping filter smooths out the response to a sudden change in level.
  • Page 107: Fail-Safe Timer

    Fail-safe Level When Fail-safe Material level is set to Value, the Fail-safe Level defines a user-defined level to report when the Fail-safe timer expires.. Fail-safe Level Values 3.6 to 22.6 mA 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 101...
  • Page 108: Process Pressure/Temperature Derating Curves

    . In the event the instrument package is replaced, the Process Device Tag shall be transferred to the replacement unit. • SITRANS LR260 units are hydrostatically tested, meeting or exceeding the requirements of the ASME Boiler and Pressure Vessel Code and the European Pressure Equipment Directive.
  • Page 109: Horn Antenna

    Tag showing one of this series, and 25589 stamped on flange. WARNING: Never attempt to loosen, remove or disassemble process connection or instrument housing while vessel contents are under pressure. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 103...
  • Page 110: Loop Power

    LR260 loop current IL Allowable operating area of SITRANS LR260 Loop Voltage versus Loop Resistance Power Supply Requirements 1000 (30.0v,845 ohms) RL=44.6*VL - 493.4 (17.8v, 300 ohms) ALLOWABLE RL=272.7*VL - 4554.1...
  • Page 111: Startup Behavior

    Startup Behavior Notes: • SITRANS LR260 is designed to start reliably with a power supply capable of delivering at least 25 mA. • When connected to a power supply with a current limit of < 25 mA, the LR260 may not start reliably.
  • Page 112 Page 106 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 113: Appendix E: Application Example

    High Calibration Pt.= 0.5 m from sensor reference point. Max.fill/empty rate = 0.1 m/ min. level In the event of a loss of echo, SITRANS LR260 is to go into Fail-safe High after 2 minutes. Calibration Point 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL...
  • Page 114 2.4.1. Fail-safe Timer 2 minutes Independent 2.4.2. Fail-safe Material Failsafe level set to parameters Level (FAILSAFE MAT. LE) High Return to Measurement: press Mode to start normal operation. Page 108 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 115: Appendix F: Hart Communications

    Device Description for the instrument in question. HART DDs are controlled by the HART Communication Foundation. Please check with the HART Communication Foundation for the availability of the HART EDD for SITRANS LR260. Older versions of the library will have to be updated in order to use all the features of SITRANS LR260.
  • Page 116: Hart Communicator 375 Menu Structure

    HART Communicator 375 Menu Structure HART Communicator 375 is supported by SITRANS LR260 HART. The menu structure Note: is aligned with the menu structure of SIMATIC PDM. SITRANS LR260 Identification Description Message Installation Date Device order number Setup Device Firmware Revision...
  • Page 117 Total Expected Sensor Life Service LCD Fast Mode LCD Contrast Service Interval Time Elapsed Since Last Service Remaining Lifetime Maintenance Required Limit Maintenance Demanded Limit Maintenance Alert Activation Service Acknowledge Total Service Interval 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 111...
  • Page 118 Calibration Interval Time Elapsed Since Last Calibration Remaining Lifetime Maintenance Required Limit Maintenance Demanded Limit Maintenance Alert Activation Calibration Acknowledge Total Calibration Interval Communication Communication control Security Lock Unlock value Language Language Page 112 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 119: Supported Hart Commands

    Supported HART Commands SITRANS LR260 conforms to HART rev. 5 and supports the following: Universal Commands 0, 1, 2, 3, 6, 7, 8, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22 Common Practice Commands 33, 34, 35, 36, 37, 38, 40,41, 42, 44, 45, 46, 48, 50, 51, 53, 54, 59...
  • Page 120: Appendix G: Firmware Revision History

    Appendix G: Firmware Revision History Firmware EDD Rev. Date Changes Rev. 1.00.01 1.00.04 Dec. 5, 2007 • Initial release Page 114 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 121: Glossary

    (decibel): a unit used to measure the amplitude of signals. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 115...
  • Page 122 1 GHz to 300 GHz. Many conductive liquids/electrolytes exhibit dielectric properties; the relative dielec- tric constant of water is 80. Page 116 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 123 Light speed is a constant 299, 792, 458 meters per second. stillpipe: a pipe that is mounted inside a vessel parallel to the vessel wall, and is open to the vessel at the bottom. 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 117...
  • Page 124 TVT (time varying threshold): a time-varying curve that determines the threshold level above which echoes are determined to be valid. waveguide antenna: a hollow, metallic tube that transmits a microwave signal to the product target. Page 118 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 125: Index

    30 power supply requirements 22 key functions 31 Loss of Echo (LOE) enclosure explanation 100 opening 20 mA output in Measure mode 99 factory defaults 7ML19985KE03 SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 119...
  • Page 126 PDM 47 settings adjust parameters via LUI 30 adjust parameters via PDM 36 SIMATIC PDM 36 features of rev. 6.0 36 Quick Start 38 simulate analog output 48 Page 120 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 127: Lcd Menu Structure

    LCD menu structure SITRANS LR260 (HART) – INSTRUCTION MANUAL Page 121 SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 128 2.2.4.2.6 WINDOW 2.2.4.3FILTERING 2.2.4.3.2 DAMPING FILTER 2.2.4.3.6 NARROW ECHO FIL. 2.2.4.3.7 REFORM ECHO 2.2.4.5 NOISE 2.2.4.5.1 CONFIDENCE 2.2.4.5.2 STRENGTH 2.2.4.5.3 NOISE AVERAGE 2.2.5 TVT SETUP Page 122 SITRANS LR260 (HART) – INSTRUCTION MANUAL SITRANS LR260 (HART) – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 129 4.17.4 MAINT DEM LIMIT 4. 1 7.5 ALERT ACTIVATION 4.17.6 TOTAL CALIB.INTRV 4. 1 7.7 MAINT STAT 4.17.8 ACK STATUS 4. 1 7.9 ACK COMMUNICATION DEVICE ADDRESS 5.2 COMM. CONTROL SECURITY LOCK 6.2 UNLOCK VALUE LANGUAGE 7ML19985KE03 SITRANS LR260– INSTRUCTION MANUAL Page 123...
  • Page 130 Page 124 SITRANS LR260 – INSTRUCTION MANUAL 7ML19985KE03...
  • Page 131: For More Information

    Siemens Milltronics Process Instruments Inc. Subject to change without prior notice Industry Automation (IA) 7ML19985KE03 Rev. 3.0 1954 Technology Drive P.O. Box 4225 *7ml19985KE03* © Siemens Milltronics Process Instruments Inc. 2012 Peterborough, ON Canada K9J 7B1 Printed in Canada email: techpubs.smpi@siemens.com www.siemens.com/processautomation...

Table of Contents