Replacing Fuses - Waters 2410 Operator's Manual

Differential refractometer
Hide thumbs Also See for 2410:
Table of Contents

Advertisement

8. Switch the pump or solvent delivery system back to the water-miscible intermediate
solvent. Flush for 10 minutes.
9. Switch the pump or solvent delivery system back to the mobile phase. Flush for 5
minutes.
10. Take the 2410 differential refractometer out of Purge mode and stop the pump or solvent
delivery system.
11. Reattach the column and reequilibrate the 2410 refractometer.

6.2 Replacing Fuses

This section describes replacing fuses in the 2410 differential refractometer.
Required Materials
Flat-blade screwdriver
Identifying a Faulty Fuse
A faulty fuse usually has a smoked glass area or broken filament. If no break is visible, you can
remove the fuse and test it with an ohmmeter.
Suspect a faulty fuse if:
• The 2410 differential refractometer fails to power on.
• The display is blank.
• The fan does not turn on.
For more information on detecting the cause of a blown fuse, refer to Section 7.3, Hardware
Troubleshooting.
Caution: To avoid electrical shock, power off and unplug the 2410 refractometer before
checking the fuse. For continued protection against fire hazard, replace the fuse only with
another fuse of the same type and rating.
Procedure
To replace a fuse:
1. Power off the 2410 refractometer and disconnect the power cable from the rear panel.
2. Insert your fingers into the fuse holder slots on the rear panel of the detector, then pull
with minimum force to remove the holder from the rear panel (see Figure 6-1).
3. Remove and discard the old fuse.
Replacing Fuses
6-3
6

Advertisement

Table of Contents
loading

Table of Contents