1-Port Calibration Procedure - Agilent Technologies U3025AE10 User Manual

Hide thumbs Also See for U3025AE10:
Table of Contents

Advertisement

U3025AE10

1-Port Calibration Procedure

1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing.
Torque to 8 in-lb. For more information press the Help button, see
Figure 43
ECal at Port 1
2. Perform a 1-Port Calibration on Port 1. On the PNA, press Calibration > Calibration
Wizard.
a. If using a mechanical cal kit, select SmartCal.
b. If using an ECal module, select ECal.
3. Continue following the Cal Wizard prompts. In the "Select Calibration Ports and ECal
Module" window, press Clear All and select Port 1, then press Next.
Figure 44
1-Port Calibration
User's Guide
Operational Check
Figure
43.
Help
47