U3025AE10
Environmental Requirements
The environmental requirements of the PNA are listed in
requirements are the same as those of the E8364B/C and N5230A/C.
CAUTION
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
Table 3
PNA Operating Environment
Temperature
Operation
Storage
Measurement Calibration
Performance Verification
Pressure Altitude
(Operation or Storage)
Enclosure Protection
Environmental Tests
The U3025AE10 complies with all applicable safety and regulatory requirements for the
intended location of use and have been evaluated to assure that they are consistent with
Agilent quality and reliability goals. On the basis of that evaluation, the following
environmental tests have been deemed unnecessary and have not been performed:
temperature, humidity, shock, vibration, altitude and power line conditions.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature
within the appropriate range. Air conditioning capacity must be consistent with the BTU
ratings given in
Table
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient temperature of
the PNA and Test Set to be maintained within ± 1 °C of the ambient temperature at
calibration.
8
0 °C to 40 °C (32 °F to 104 °F)
–40 °C to +70 °C (–40 °F to +158 °F)
20 °C to 26 °C (68 °F to 79 °F)
Temperature must be within 1 °C (1.8 °F) of the temperature
at which the measurement calibration was performed.
0 to 4600 meters (~ 15,000 feet)
IP 2 0, according to IEC 529
2.
General Specifications
Table
3. Note that these
User's Guide