Service

Changing the Orientation of or Replacing the Contact Assembly

Changing the Orientation of or Replacing the Contact
Assembly
If the center electrode of the contact assembly is smeared, an unused clean area of the
center electrode can be used simply by changing the orientation of the contact assembly.
This section describes the method for changing the orientation of or replacing the contact
assembly.
To change the contact assembly orientation or replace the contact assembly, it is necessary
to use a 1.5-mm Allen wrench (Agilent part number: 8710-0909) and a Phillips
screwdriver.
Procedure for Changing the Orientation of or Replacing the Contact
Assembly
1. Take off the guide holder, and remove the device guide and electrode plate. (See to
Figure 3-4 on page 36).
2. Remove the four screws from the upper side of the fixture, and then remove the contact
assembly and flange.
Figure 6-3
Removing the Contact Assembly
74
Chapter 6