U3025AE10
5. At the Calibration Completed prompt, select Save As User Calset and type the name
999.1. Overwrite the Calset if it already exists. Press Save.
NOTE
If you do not have a key board, select Save As User Calset > Edit Name and
save as 999.x. X is the port number you are calibrating. See
the numeric keypad on the PNA front panel to enter "999.1."
Figure 51
Calibration Complete
6. Repeat
step 1
thru
finished, there should be twelve Cal Sets saved with the titles "999.1" thru "999.12"
(12-Port).
If you are using an ECal module you can verify the individual port calibration by
selecting Response > CAL > More > ECAL > ECAL Confidence Check. For further
information refer to the system Help menu.
7. Press Trace/Chan > Trace > Delete Trace. There should be no traces on the PNA screen.
User's Guide
step 5
(1-Port Calibration Procedure) for Ports 2 thru 12. When
Operational Check
Figure
51. Use
53