2. Continue following the Cal Wizard prompts. On the "Select Calibration Ports and ECal
Module" window, press Select All or select the ports you are calibrating and press Next.
Figure 23
12-Port Calibration
3. Connect the ECal or the mechanical cal kit to the ports you are calibrating following the
Cal Wizard prompts and press Measure after each connection.
4. At the Calibration Completed prompt, select Save As User Calset and type the name
desired. Press Save.
5. After calibrating test set ports, use a quality load and short to verify the calibration on
each port or end of the test cable. Measure reflection and confirm the return loss is as
expected. If the result is not as expected, repeat the calibration without the test set and
ensure that the PNA is in standard (non-multiport) mode.