10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in
page
43. This can be done using Limit Lines (press Trace > Limit Test) or Marker. The
trace should be above the limit line and PASS will be displayed on the screen. Refer to
Figure
48.
11.Repeat
step 9
and
Figure 48
E8364B/C Reflection Tracking Trace (Ports 1-12)
50
step 10
for Cal Sets "999.1" thru "999.12" (12-Port).
Operational Check
Table 12 on
User's Guide