U3025AE10
8. To launch the Cal Set Viewer toolbar. Select Response > CAL > Manage CALS >
CAL Set Viewer.
Figure 52
Calibration, Cal Set Viewer
9. From the Cal Sets drop-down menu, select 999.1 and check Enable. Select the
Reflection Tracking(1,1) term in the center drop-down menu and ensure that the
Enable and Error Terms are selected.
NOTE
You may also create a table on the network analyzer and enter the limit line
stimulus and response values. Select Limit Test ON and Limit Line ON, and
select Show Table. See
Figure 53
PNA-X Setting the Test Limits
54
Figure
53.
Operational Check
User's Guide