Cautions

CAUTION
The CAUTION notice denotes a hazard. It calls attention to an operating
procedure, practice, or the like, which if not correctly performed or adhered to,
could result in damage to the product or loss of important data. Do not
proceed beyond a CAUTION notice until the indicated conditions are fully
understood and met.
Cautions applicable to this instrument are:
CAUTION
Always use the three-prong ac power cord supplied with this instrument.
Failure to ensure adequate earth grounding (by not using this cord) can cause
instrument damage.
CAUTION
This instrument has autoranging line voltage input; be sure the supply
voltage is within the specified range.
Ventilation Requirements: When installing the instrument in a cabinet, the
CAUTION
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
CAUTION
This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 61010-1:2000, and 664 respectively.