6 .3.2 Removing Water-soluble Contamination

Parts and Tools
Procedure
1. Moisten a clean cloth with water.
. Wipe off the area with the cloth.
2
. Dry with a lint-free cloth.
3

6 .3.3 Replacing the LED modules

Burning hazard due to hot light sources
Light sources can become hot during processing.
4
4
Eye damage or skin irritation due to hazardous light emission
The light source emits LED radiation and UV radiation. Eye damage or skin irritation may result
from exposure.
4
4
4
Parts and Tools
ü
Prerequisite
ü
Procedure
1. Loosen the four screws on the rear cover.
2
. Remove the cover.
Instruction Manual ZEISS Axiovert 5 digital | en-US | Rev. 1 | 431030-7021-101
Clean cloth
Lint-free cloth
à
A mild detergent may be added to the water (no solvent!).
CAUTION
Avoid touching the hot light source housing.
Let the light source cool down before touching it.
CAUTION
Never look directly into the light-emitting aperture of the light source.
Avoid exposure of skin to the radiation. Use suitable protective equipment/protective
clothing if required.
Before installing or removing the light source always make sure it is switched off.
Hex Key, 2.0 mm
Suitable LED module
The microscope is switched off.
The microscope is unplugged from mains power.
6 Care and Maintenance | 6.3 Maintenance Work
39