6 Care and Maintenance | 6.2 Maintenance Schedule
6 .2 Maintenance Schedule
The recommended intervals for maintenance depend on the total uptime of the microscope.
Interval
daily
Tab. 2: Maintenance Plan
6 .3 Maintenance Work
6 .3.1 Cleaning an Optical Surface
NOTICE
Damage of optical surfaces due to improper cleaning
4
4
4
Parts and Tools
Procedure
1. Moisten a cotton swab or a clean cloth with an optical cleaning solution, if necessary.
2 . Wipe optical surfaces in a circular motion to-
3 . Dry with a lint-free cloth.
3 8
Part/Component
-
Remove dust from the optical surface slowly and carefully.
Remove dust on optical surfaces with a natural-hair brush or blow it off with a rubber bel-
lows.
Avoid touching optical surfaces with fingers.
Clean cloth
Cotton swab
Optical cleaning solution (85% n-hexan and 15 vol% isopropyl alcohol (IPA))
Lint-free cloth
wards the edge of the optics with slight pres-
sure.
Activity
Check the power cable and each plug for
possible damage.
If any damage is observed, turn the instru-
ment off and secure it against inadvertent
restarts immediately. Call in a qualified pro-
fessional to remedy the problem.