Tests and Adjustments
Adjustments
2. Follow the instructions and prompts as they are displayed.

Receiver Adjustment

The receiver calibration is used to adjust the network analyzer receivers for a
flat response across its full frequency range:
1. A power meter/sensor is connected to Port 1, as shown in
2. A cable is inserted between the power sensor and the test port, as shown
3. The same cable is connected between test port 1 and test port 2, as
Data obtained during this adjustment are stored in the mxcalfile_pxx files in
flash memory on the test set motherboard, with a backup copy stored on the
hard disk drive. The data are used in subsequent measurements.
Solid state drives can be swapped or replaced without concern for the
mxcalfile_pxx files. If the test set motherboard is replaced, the PNA firmware
will automatically create new primary mxcalfile_pxx files from the backup
copies on the hard drive.
These files can be recreated by performing another receiver calibration
adjustment.
Equipment Used for the Receiver Adjustment
Equipment Type
Power meter
Power sensor, 2.4 mm
RF Cable, 2.4 mm (f) to 2.4mm (f)
Adapters may be required, depending on the equipment you use.
Procedure
1. Connect the equipment as shown in
Keysight N5244B/45B Service Guide
establish a reference for flatness.
in
Figure
3-18, to establish a reference for the cable.
shown in
Figure
3-19, and a signal from Port 1 is used to adjust the "B"
receiver at Port 2.
The adjustment is repeated using a signal from Port 2 to adjust the "A"
receiver at Port 1.
between the network analyzer and the power meter.