Tests and Adjustments
Adjustments
Figure 3-20
PNA Noise Adjustment Dialog Box
3. Click Begin.
4. As shown in
Keysight N5244B/45B Service Guide
— Set the GPIB address to match the address of the power
meter.
— If using a single channel power meter, use the default setting
Ch. A. If using a dual channel power meter and the sensor is
attached to the B channel, select Ch. B.
— If the sensor has already been zeroed and calibrated, select
Skip Zero and Cal before first use.
Figure
3-21, connect an RF cable between the network
analyzer test ports 1 and 2.
3-59