Keysight Technologies N5244/5B Service Manual page 147

2-port and 4-port pna-x microwave network analyzers
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
2. Unplug the power to the analyzer and disconnect all front and rear panel
connections except installed jumpers.
3. Remove the outer and inner covers from the analyzer. Refer to
the Covers" on page
Procedures described in this document are performed with power
supplied to the product while protective covers are removed. Energy
available at many points may, if contacted, result in personal injury.
4. With the covers off, plug in the analyzer and turn on the power.
Do not operate the analyzer with the outer cover removed for more than
30 minutes, as this could cause the analyzer to overheat which could
result in costly damage.
Checking the A27 and A28 Mixer Brick Receiver Outputs
Set the network analyzer for an S
GHz.
1. Locate the following flexible cables at the receiver IF outputs of the mixer
brick(s).
2. Disconnect the flexible cable at the suspect receiver.
3. Connect the spectrum analyzer to the suspect receiver connector.
Keysight N5244B/45B Service Guide
7-8.
— 2-port models
— Receiver A; W61 of A27
— Receiver R1; W62 of A27
— Receiver R2; W63 of A27
— Receiver B; W64 of A27
— 4-port models
— Receiver A; W61 of A27
— Receiver R1; W62 of A27
— Receiver R2; W63 of A27
— Receiver B; W64 of A27
— Receiver C; W68 of A28
— Receiver R3; W67 of A28
— Receiver R4; W66 of A28
— Receiver D; W65 of A28
measurement with a CW frequency of 1
11
"Removing
4-53

Advertisement

Table of Contents
loading

Table of Contents