Safety and Regulatory Information
Regulatory Information
Environmental Information
Samples of this product have been type-tested in accordance with the
Keysight Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation and end-use; those
stresses include but are not limited to temperature, humidity, shock,
vibration, altitude, and power-line conditions. Test methods are aligned
with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.
Parameter
Operating Environment
Operating Temperature
Storage Temperature
Error-corrected range
Keysight N5244B/45B Service Guide
The crossed-out wheeled bin symbol indicates that separate collection for
waste electric and electronic equipment (WEEE) is required, as obligated by the
EU DIRECTIVE and other National legislation. Please refer to
www.keysight.com/go/takeback
Keysight, in addition to product takeback instructions.
China Restricted Substance Product Label. The EPUP (environmental protection use
period) number in the center indicates the time period during which no hazardous or
toxic substances or elements are expected to leak or deteriorate during normal use
and generally reflects the expected useful life of the product.
Universal recycling symbol. This symbol indicates compliance with the China
standard GB 18455-2001 as required by the China RoHS regulations for
paper/fiberboard packaging.
This mark indicates product has been designed to meet the requirements of "IP x y",
where "x" is the solid particle protection and "y" is the liquid ingress protection.
Required Values/Ranges
For Indoor Use ONLY
This instrument has no air filters on the fan ports. Operation in dirty,
dusty, or similar environments must be avoided.
0 °C to +40 °C
The instrument powers-up and displays no error messages within this
temperature range (except for "source unleveled" error message that may
occur at temperatures outside the specified performance temperature
range of 25 ± 5 °C).
–40 °C to +70 °C
23 °C ±3 °C with less than 1 °C deviation from calibration temperature.