Keysight Technologies N5244/5B Service Manual page 651

2-port and 4-port pna-x microwave network analyzers
Table of Contents

Advertisement

General Purpose Maintenance Procedures
Error Terms
Figure 8-3
Flowgraph of Three-Port Error Terms
Keysight N5244B/45B Service Guide
where:
E = error term
DIR = Directivity
MAT = Forward Source Match and Reverse Load Match
TRK = Forward Reflection Tracking and Reverse Transmission Tracking
For the case of a full 3-port calibration, port 1 has
three Match error terms:
S11 source match
S12 load match
S13 load match
and three Tracking error terms:
S11 reflection tracking
S12 transmission tracking
S13 transmission tracking
There are six isolation terms not shown.
8-5