6 Operation Instructions
3. Press
or
tion or select the data stored cycle.
Function "OFF" = [off]
Data stored every 2 min. = [2Min]
Data stored every 5 min. = [5min]
4. Press
button to return to "Function Selec-
tion Mode".
NOTE:
The maximum writable limit of the memory device is 1 million cycles. If the sensor is operated 24 hours
per day, the durability is calculated as below:
5 minutes x 1 million cycles = 5 million minutes = 9.5 years
2 minutes x 1 million cycles = 2 million minutes = 3.8 years
6.11   Flow Sensor Display Mode Selection
[F-10] Select to display Instantaneous Flow or Accumulated Flow Mode.
1. Press
or
tion Mode" to display [F-10] [ di5].
2. Press
button to switch to "Display Mode
Selection".
3. Press
or
mode.
Instantaneous Flow Mode = [in5]
Accumulated Flow Mode = [add]
NOTE: When the measured flow rate range is
500mL/min or 1000mL/min, the accumulated
flow is selected and the unit will become in
mL/min.
4. Press
button to return to "Function Selec-
tion Mode".
6.12   Synchronizing the Flow Value Between the Analog Output and the Display
[F-80] Select to synchronize the flow value between the analog output and the display.
This function is available only for the flow output.
1. In function selection mode, press
to display [F-80] [5yn].
2. Press the
button to switch to the display
synchronization.
34 / 46
button to turn off the func-
button at "Function Selec-
button to select display
or
EN-US · 30.30.01.02533 · 02 · 12/22