STX START-UP PROCEDURE for L PLUS CONTROL
1>
Turn on the air compressor.
2>
Open the premix laser gas cylinder until the valve is seated in the open position.
Primary Gauge:
Secondary Gauge:
3>
Open the assist gas cylinders until the valve is seated in the open position.
IMPORTANT: OXYGEN CAN ONLY BE USED IN THE PORT LABELED "O2 GAS IN".
DO NOT CONNECT OXYGEN TO THE "3
OXYGEN SECONDARY PRESSURE:
4>
Turn on the main breaker.
NOTE: The main breaker should be left on at all times in order to keep the resonator
and/or refrigerant compressor temperature stable.
The following buttons/keys are located on the operating panel.
5>
Press the "POWER ON" button.
6>
Press the HOME key. ("HOME")
7>
Turn the "LASER" key switch to the "ENABLE" position. ("1" = ON)
8>
Press the "LASER POWER" ON key.
9>
When the "LASER POWER ON" indicator LED stops flashing and remains on
continuously, press the "LASER HIGH VOLT" ON key.
10>
When the "LASER HIGH VOLT" indicator LED stops flashing and remains on
continuously, the machine is ready for operation.
NEVER REMOVE ANY COVERS FROM THE LASER POWER UNIT OR LASER
RESONATOR WHILE LASER POWER IS ON DUE TO EXTREMELY HIGH VOLTAGES
PRESENT DURING THE HIGH VOLT ON STATE.
>142 psi.
35 - 40 psi.
RD
ASSIST GAS" PORT.
140 PSI. (NO GREATER)
(IT WILL TAKE ABOUT 30 SEC. FOR NC ON)
( "1"=ON, "0"=OFF )
** IMPORTANT **
( 1=ON
0=OFF )
4