Attaching The Marker - Siemens MAGNETOM Avanto Operator's Manual

4 ch bi breast / 16 ch ai breast
Hide thumbs Also See for MAGNETOM Avanto:
Table of Contents

Advertisement

5
Biopsy examination
Compressing the breast
40
2
For lateral access: If the lesion is located near the axillary tail, use
straps at the elbow to elevate the arm on the puncture side.
3
For cranio-caudal access: Position the arm on the frame next to the
body for easy access to lesions in the vicinity of the axillary tail or
the breast wall.
4
For medial access: Slide the insertion plate for the recess laterally
into the frame so that the other breast does not restrict free access
to the lesion.
5
If necessary, optimize the height of the horizontal or vertical slat
plates using the screws.
 CAUTION
Overcompression!
Pressure leading to injury of the breast
Incorrect diagnosis due to insufficient contrast
enhancement
Apply just enough compression to immobilize the breast, but
do not cause pain to the female patient and ensure that
contrast enhancement is continued correctly.
1
Slide the fixation unit with the horizontal or vertical slat plates
carefully together to compress the area of the breast to be
examined.
2
Adjust compression to stabilize the breast during needle insertion,
but avoid overcompression.

Attaching the marker

5.2.4
The marker may only be used for establishing the reference position
for needle insertion.
Marker is disinfected.
Fill the marker with a syringe (e.g., with 1:200 diluted 0.5 molar
contrast medium). Ensure that the marker is filled completely.
4 Ch BI Breast / 16 Ch AI Breast | Operator Manual
Print No. MR-04010G.623.22.02.02