General safety information
12
When moving the patient table, the coordinates computed for
the breast biopsy are deleted. However, it is still possible to print
out the worksheet.
◾
Print out the worksheet directly after the offset coordinates are
generated.
◾
Compare the data on the print-out with those on-screen. The
data have to match.
2.2
Risks related to incorrect handling of the coil
CAUTION
Damaged biopsy device!
Risk of injury
◆
Prior to each use, check all components and individual parts
of the biopsy device for damage.
◆
Use only undamaged components.
CAUTION
Breast implants!
Risk of rupture
Risk of damage caused by puncture
◆
Use the greatest of care when performing biopsies on female
patients with breast implants.
If a previously enhanced lesion is not visible during the
examination using the breast biopsy device, perform a follow-up
examination the next day without applying compression.