Biopsy examination
Background information:
Breast Biopsy (dialog window)
50
1
Move the patient table out of the magnet.
2
Position the wire anchor of the marking wire in the lesion, and
after that remove the core needle device from the breast.
3
Verify the correct wire position by performing another
measurement using the 2D protocol.
The Breast Biopsy dialog window is divided into two areas:
Switching field: here you'll edit the reference images, set markers,
and calculate the data for the biopsy.
Data field: here the text fields are used to display the current
patient data (name, date of birth) and the data calculated on the
basis of the angulation used.
(1) Activate/deactivate the crosshair
(2) Zoom/pan images
(3) Modify graphic
(4) Store the marker position
(5) Store the lesion position
(6) Image comment
(7) Calculate/save target coordinates
(8) Print out a worksheet (anonymously, if necessary)
The functions for graphic slice positioning are not active while
the Breast Biopsy dialog window is open.
4 Ch BI Breast / 16 Ch AI Breast | Operator Manual
Print No. MR-04010G.623.22.02.02