Manual
00809-0100-4891
Note
A. Except for JP5, JP7, and JP8 on the input/output (IO) board, jumper and switch settings are factory set and are
shown for reference only.
B. IO board: 4-20 mA/HART loop power settings
•
JP5
—
Pins 1-2: internal power Rosemount 6888Xi to Rosemount 6888 analyzer
—
Pins 2-5: external power Rosemount 6888Xi to Rosemount 6888 analyzer (requires 2,500 resistor across
J4, PR+ to PR-)
•
JP7/JP8
—
Pins 1-2: internal power to DCS
—
Pins 2-3: external power Rosemount 6888Xi to DCS
Note
IO board Channel 2 is a duplicate of Channel 1.
4.7
Connect the analyzer probe with integral
autocalibration to F
communications
This probe contains gas-switching solenoids enabling the Rosemount 6888 Xi
Electronics, Field Communicator, or Asset Management Solutions (AMS) software
to control the introduction of calibration gases.
You can initiate calibrations in the following ways
•
Automatically with a calibration recommended diagnostic
•
Automatically with time since last calibration
•
Manually with external dry contact
•
With the Rosemount 6888 Xi local operator interface (LOI). You can only use
integral autocalibration with a Rosemount 6888 Xi.
Procedure
1. Remove the two covers from the analyzer.
2. Connect the line (L1 wire) to the L1 terminal, the neutral (L2) wire to the
3. Connect the F
Manual
L2/N terminal, and the ground wire to the ground lug.
The analyzer accepts line voltage at 120/240 VAC ±10 percent, 50/60 Hz. No
setup is required.
OUNDATION
the FF segment.
OUNDATION
Fieldbus wires from the analyzer side housing to
Install
January 2023
Fieldbus
™
35