Write Commands - Lauda LOOP 100 Operation Manual

Thermo-electric circulation thermostat
Table of Contents

Advertisement

5.11.3

Write commands

Command
OUT_SP_00_XXX.XX
OUT_SP_04_XXX
OUT_SP_05_XXX
OUT_PAR_00_XX.X
OUT_PAR_01_XXX
OUT_PAR_02_XXX
OUT_PAR_03_XX.X
OUT_MODE_00_X
START
STOP
Table 5: Acceptable data formats
-XXX.XX
-XXX.X
-XX.XX
-XX.X
-X.XX
-X.X
-.XX
-.X
30 / 41
The command from the computer must be made with a CR, CRLF, or
LFCR.
CR = Carriage Return (Hex: 0D); LF = Line Feed (Hex: 0A)
The response from the thermostat is always made with a CRLF.
After each command sent to the thermostat, it is necessary to wait for
the reply before sending another command. This ensures that the
sequencing of inquiries and answers is clear.
Table 4: Example for set value transfer from 30.5 °C to the thermostat.
Computer
"OUT_SP_00_30.5"CRLF
Write commands are data specifications from the PC to the thermostat.
Description
Desired value transfer with maximum 3 places in front of the decimal point and
a maximum of 2 places after.
Write [Hi] Outflow temperature limit upper value
Write [Lo] Outflow temperature limit lower value
Configuration of the control parameter Xp
Configuration of the control parameter Tn (5 - 180 s; 181 = Off)
Configuration of the control parameter Tv
Configuration of the control parameter Td
Keyboard master: 0 = free / 1 = blocked (corresponds to: "KEY")
Turns the device on (from standby).
Switches the device to standby (switches the pump and Peltier elements off).
Take the following information into account:
" " (blank space) can also be used for "_".
Response from thermostat will be "OK" or, if there is an error, "ERR_X".
-XXX.
-XXX
-XX.
-XX
-X.
-X
.XX
.X
LOOP
Thermostat
"OK"CRLF
XXX.XX
XXX.X
XX.XX
XX.X
X.XX
X.X
XXX.
XXX
XX.
XX
X.
X
V3