Lauda Heat Transfer Liquid - Lauda LOOP 100 Operation Manual

Thermo-electric circulation thermostat
Table of Contents

Advertisement

4.2

LAUDA heat transfer liquid

Table 3: Approved heat transfer liquids
LAUDA
Chemical
designation
name
Decalcified
Aqua 90
water
Monoethylene
Kryo 30
glycol/water
mixture
Heat transfer liquid Kryo 30
Heat transfer liquid, water
16 / 41
Note the following instructions:
The heat transfer liquids each cover a recommended temperature range
and must be suitable for the temperature range associated with their
application.
Never use contaminated or degenerated heat transfer liquids.
You can request a copy of the safety data sheets for the heat transfer
liquid at any time, if necessary.
Viscosity (kin) in
Temperature
mm²/s at
range in °C
(at 20 °C)
5 - 90
1
4
-30 - 90
The water content decreases during longer periods of operation at
higher temperatures and the mixture becomes combustible (flash point
119 °C). Check the mixing ratio using a hydrometer.
The proportion of alkaline earth ions in the water must be between
0.71 mmol/L and 1.42 mmol/L (equivalent of 4.0 and 8.0 °dH). Harder
water leaves limescale deposits in the device.
The pH value of the water must be between 6.0 and 8.5.
Due to their corrosive properties, distilled, deionized, fully desalinated
(FD) water, or ocean water cannot be used. High-purity water and dis-
tillates are suitable as a medium after 0.1 g of soda (Na
carbonate) is added for every liter of water.
Avoid chlorine content in the water at all costs. Do not add chlorine to
the water. Cleaning agents and disinfectants, for example, contain
chlorine.
The water must be free of impurities. Water with iron content is unsuit-
able due to rust formation and untreated river water is unsuitable due to
algae growth.
The addition of ammonia is not permitted.
LOOP
Viscosity (kin) in
mm²/s at
temperature
5 L
LZB 120
¾
50 at -25 °C
LZB 109
Container size
Cat. No.
10 L
20 L
LZB 220
LZB 320
LZB 209
LZB 309
CO
, sodium
2
3
V3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Loop 250L100L250