Nonin 8600V Operator's Manual page 13

Veterinary pulse oximeter
Table of Contents

Advertisement

Audible Alarm Enable/Disable
Momentarily pressing the
override the 2-minute period, re-enabling the audible alarm.
NOTE:
Pressing this button has no effect if the audible alarm has been permanently disabled.
Calibrate Recorders
When using a strip-chart recorder for outputting data, calibration between the recorder and the 8600V may be
necessary. To do this during normal operation, continuously hold the
activate the recording output calibrate sequence. The user can select the calibration sequence with Option Switch
3.
NOTE:
The calibration signals are sent via the SpO
line.
Standard Calibration:
The value sent for the SpO
seconds) and then repeat. The value sent for the pulse rate will alternate from 0 BPM to 125 BPM to 250
BPM and then repeat. To end calibration sequence, release the
return to normal operation.
Alternate Calibration:
The value sent for the SpO
AUDIO
button beginning at 0% through 100% and then repeat. The value sent for the pulse rate will
remain at 60 BPM. To end calibration sequence, hold the
will return to normal operation.
AUDIO
button disables the audible alarm for 2 minutes. Pressing this button again will
will alternate from 0% (for 5 seconds) to 100% (for 6 seconds) to 50% (for 5
2
will step through at 5% intervals when actuated by the user by pressing the
2
AUDIO
button more than 5 seconds to
and pulse rate lines, not the serial output
2
AUDIO
AUDIO
button for 5 seconds, and the 8600V
10
button, and the 8600V will