SpO 2 Low Limit
The button labeled with the
the SpO
low alarm are given in Table III.
2
Press the
button to the left of the number to decrease value.
Press the
button to the right of the number to increase value.
Pulse Rate High Limit
The button labeled with the
the pulse rate high alarm are given in Table III.
Press the
button to the left of the number to decrease value.
Press the
button to the right of the number to increase value.
Pulse Rate Low Limit
The button labeled with the
the pulse rate low alarm are given in Table III.
Press the
button to the left of the number to decrease value.
Press the
button to the right of the number to increase value.
Pulse Volume
The 8600V may be set such that it will emit a short tone on each pulse detected. Turn the pulse volume control,
marked with the
SpO
level changes. The pitch is lower in frequency with lower SpO
2
the unit loses track of the pulse signal.
Alarm Volume
Adjust the alarm volume intensity by rotating the thumbwheel volume control on the front of the 8600V marked with
the
symbol. The OFF position is only effective when option switch 1 is in the up position. The option
switches are located on the rear of the 8600V.
NOTE:
NONIN does not recommend disabling the audible alarm when monitoring in critical
situations.
To generate an alarm for test purposes, momentarily disconnect the patient cable assembly from the monitor. Press the
latches and pull the connector out. This will generate a sensor alarm. The
audible alarm will sound provided the alarm volume control is not disabled, and option switch 1 is not in the up
position, and the two-minute audible disable period is not active.
next to the SpO
2
next to the ♥ display sets the high pulse rate alarm limit. The available values for
next to the ♥ display sets the low pulse rate alarm limit. The available values for
symbol, until an audible pulse is heard with each beat. The tone will change pitch as the
% display sets the low SpO
levels. A noticeably higher pitched tone is heard if
2
SENSOR
16
alarm limit. The available values for
2
indicator will illuminate. The

Advertisement

Table of Contents
loading

This manual is also suitable for:

8600mv

Table of Contents