Communications

The 8600V may communicate in two forms: serial and analog communications. It may be connected to three different
devices as given in Table IV. Real-time data may be output to all listed devices from a 8600V or 8600MV, and stored
memory data may be output to all listed devices from a 8600MV.
Analog
Serial
Communication
Strip Chart
Link
Recorder
- Real-time
X
- Memory Output
- Real-time
- Memory Output
Table IV: Communications Links
Pin
Number
Signal
1
Ground
2
+5 V
3
Analog/Serial
Control
4
Not Connected
5
Not Connected
6
Alarm
7
Pulse
8
Ground
9
Serial Input
10
Serial Output
11
Not Connected
12
SpO 2
13
Pulse
OR
Rate
14
Not Connected
15
+5 V
Table V: Recording Output Interface Assignments
flashing
indicator
green
real-time data
transfer
yellow
waiting for data
red
real-time error
Table VI: Status Indicator During Data Transfer
8586PI
PC
Printer Interface
(nVISION software)
X
Assignment
System Ground
External +5 VDC Supply
Set high to send (high speed) serial data
Spare Input Line
Spare Output Line
+5 V Activated with Audible Alarm
+5 V Activated with each Pulse
(corresponds to perfusion indicator)
System Ground
0 ± 5 V levels
0 ± 5 V levels
Not Connected
0 to +1 V levels corresponding to 0 to
100% SpO 2
Plethysmographic pulse waveform OR
0 to +1 V levels corresponding to 0 to
250 pulse per minute
Not Connected
+5 VDC Internal Supply