Nonin 8600V Operator's Manual page 26

Veterinary pulse oximeter
Table of Contents

Advertisement

Memory Option (8600MV Only)
The memory option is identified by the "M" in the model number (i.e. 8600MV as opposed to the 8600V). The
8600MV Pulse Oximeter can collect and store up to eighteen hours of Patient SpO
information may be output via the 15-pin sub-D connector labeled as "RECORDING OUTPUT" on the rear of the
8600V.
The solid-state memory in the 8600MV functions much like an "endless loop." When the memory fills up, the unit
begins overwriting the oldest locations with the latest data.
Each time the 8600MV is powered up, the current time/date information (if the clock is set properly) is stored in
memory to allow quick differentiation of recording sessions. Patient SpO
four seconds. The stored resolution of the oxygen saturation is in 1% increments in the range of 0 to 100%. The
stored pulse rate ranges from 18 to 300 BPM. The stored values have a resolution of 1 BPM from 18 to 200 and a
resolution of 2 BPM from 201 to 300.
During the data printing, the last data recorded is the first data printed. During data output, the
the status of the data transfer.
Recording Sessions
Each time the 8600MV is turned on, data is automatically collected.
NOTE:
Only recording sessions longer than one minute are kept in memory.
Memory Output Mode
To output the data stored in the memory of the 8600MV, start with the unit OFF and then:
AUDIO
1) Hold the
AUDIO
2) Release the
3) Observe the hour and minute of the output data will be displayed in the SpO
4) Data is automatically transferred from the memory.
5) When data is done being output,
6) Restart the 8600MV.
Data is transferred at a rate of 20 minutes of collected data per second. An 18-hour recording session (the
maximum memory saved) is transferred in approximately 1 minute. After all the data is transferred, the 8600MV
should be shut off prior to collecting new patient data. Outputting the memory does not clear any data from the
memory. The patient information is held in memory until it is overwritten by new data unless you clear the
memory by pressing the
Mode.
NOTE:
If the memory has not been cleared, the 8600MV may be downloaded multiple times with
the same data.
button while pressing the "" switch.
button when
MEMORY
CLEAR
button for a minimum of 5 seconds while the 8600MV is in the Memory Output
and pulse rate information. This
2
and pulse rate are sampled and stored every
2
and ♥ LEDs.
is displayed in the SpO
2
is displayed.
23
indicator displays
and ♥ LEDs.
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

8600mv

Table of Contents