Scaleo medical Poweo 300 User Manual page 38

Patient lifters
Table of Contents

Advertisement

The aim of adding this part is to reduce the minimum height of the hoist, making it easier to lift
patients from the floor.
Other specifications:
Classification
Maximum load
Materials
Lifting actuator
Front wheels (EN 12531 compliant)
Rear wheels ( EN 12531 compliant)
Number of lifts with a 100% charged
battery
Condition of use
IP classes
Service factor
Compliance with standards
Temperature of use
Sound power level
Technical specifications of the accessories
Technical specification of the S16 30 200 battery:
Technology
Plastic case
Voltage
Capacity
IP class
Compliance with standards
Temperature range of use
Condition of use
Security devices
DOMU124_EN_v2_20221206_Poweo® 300 User manual
Class I medical devices (European Regulation 2017/745)
300 kg
Steel and aluminum
24V – 10A
2 x 75 mm
2 x 100 mm with brake
More than de 50
Indoor use only
IPX4 in use, IPXX in charge
2/10
EN ISO 10535:2007
+10°C / +40°C
L wa = 49 dB
NiMh
ABS PC UL94V0
24V
3 Ah
IPX4 in use, IPXX in charge
EN 60601-1, UL94V0 case
+10°C / +40°C
Indoor use only
- Protection against explosion risks: the case is vented
for removal of exhaust gas
- Built-in electronic protection against surges on
intensity, short-circuit ...