Scaleo medical Poweo 300 User Manual page 39

Patient lifters
Table of Contents

Advertisement

Technical specifications of the S17 50 001 battery charger:
The charger provided is able to work with any AC power value: voltage from 100V to 240V,
frequency 50Hz or 60Hz.
The only adaptation needed is the AC cord.
Before ordering, please inform our services of the place of use: SCALEO Medical will provide the
good model of the AC cord.
Input AC voltage
Input AC frequency
Output DC voltage and current
Charging capacity
Maximum output current
Compliance with standards
Temperature range of use
Condition of use
Security devices
Usage
(1)
The charge of NiMh batteries is a smart charge: a complete charge consists in a fast charge
period, then a slow charge period, then a compensation charge.
Optional accessories (Scalis® scale, electric tilting suspension, slings) are delivered with their
own user manuals including complete technical specifications.
39
- Built-in electronic protection against overheating
during charging
100-240VAC
50-60Hz
N.A. depends on the charge cycle
NiMh batteries, 10 to 20 cells
0.9A
EN ISO 60601-1 , EN ISO 60601-1-2, UL94V0 case
-10°C / +40°C
Indoor use only
Operating voltage LED
-
-
Modulation of the charging current according
to the state of charge of the battery
Automatic load stop with visual alarm in case
-
of overheating of the battery during charging
As a « table » charger or attached to the wall
(pins and screws provided)
DOMU124_EN_v2_20221206_Poweo® 300 User manual
(1)

Advertisement

Table of Contents
loading