Periodic Inspection And Preventive Maintenance; Periodic Control/ Preventive Maintenance Checklist - Scaleo medical Poweo 300 User Manual

Patient lifters
Table of Contents

Advertisement

Periodic inspection and preventive
maintenance
The aim of a periodic control visit is to control the safety and the maintenance issues on the
equipment. This control, as well as the preventive maintenance must be performed regularly every
year for the patient lifter and twice a year for the slings – according to annex B of the EN ISO 10535
standard and the recommendations from health authorities.

Periodic control/ Preventive maintenance checklist

Ref : FOR_06_22
Version: A
Periodic inspection :
CLIENT
Name :
Poweo® model: 140
150
200
215
300
2- Checkpoints
PI
PM
Completed actions
Cleaning / Disinfection
Maintenance indicator :
Off
Resetting the maintenance program
Saving data
Base :
Base geometry and flatness
Fixation of the base with the column
Cleaning and lubrication of wheels (front &
back)
Electric emergency lowering
Electric opening of the base
31
Form attached to:
PRO_06_01
PERIODIC INSPECTION OF THE POWEO®
With electric opening of the base
Preventive maintenance:
SAGE Client N°:
City/Department :
180
185
Serial N°:
Orange
Red
DOMU124_EN_v2_20221206_Poweo® 300 User manual
Written by : CASTINEIRA
Gilbert
OK
Observations
Date :
22/01/2020