Instructions For Use - Scaleo medical Poweo 300 User Manual

Patient lifters
Table of Contents

Advertisement

Instructions for use

Security devices
Emergency stop:
®
The Poweo
patient lifters are equipped with an emergency stop button that will
turn off the lifter when pressed. The emergency stop button is located on the
top of the battery. Once pressed It must be turned clockwise to release it
Emergency lowering devices:
If the battery is discharged during a lift, there is at least enough power to lower
the patient down again.
In case the remote control does not operates correctly, an emergency electric
lowering is still possible by activating the hidden button located behind the yellow
label of the base.
Perforate the center of the yellow sticker and press the hidden button with a small screwdriver and
the patient lifter will perform an emergency lowering. Maintain the pressure on the button to
obtain the desired height.
In case of a complete failure of the electronics, a manual emergency lowering is
available.Activate the RED lever/trigger on the actuator, by pulling it upward, and
maintain the position until the actuator reaches the desired height.
Lowering speed will depend on the charge and the pressure on the trigger.
According to the ISO EN 10535 standard, the Poweo
with a patient weighing more than its maximum safety load capacity. For the Scalis®
equipped version, the scale will display a "> MAX!" message to indicate that the patient
lifter is overloaded.
The opening of the legs is also secured: in the case of a contact between a leg and an
obstacle, the legs will automatically stop moving. The leg opening movement stop is
DOMU124_EN_v2_20221206_Poweo® 300 User manual
®