Upon request, the charger's power connector can be delivered protected by a plug that
ensures additional protection of the charger against splashing water. Remove this plug for
charging, and be sure to replace it after recharging.
To charge the patient lifter, use only the supplied cable: charging a battery with an
unsuitable cable can cause the cable and/or battery to overheat and damage all
components, or cause a fire.
After each charge, be careful to unplug the power cord without pulling it to avoid
damaging the cord.
It is recommended to have 2 batteries to replace the discharged battery with the charged
one. Therefore, Poweo® will remain functional at all times.
Using the external charger
The external charger allows you to charge one battery while a second battery in being used.
The removal of the SCALEO® battery from the external charger does not require any manipulation.
It simply needs to be removed by pulling it towards you. Its shape prevents it from being
accidentally connected backwards or improperly, either in the patient lifter or in the external
charger.
The LED indicator of the charger displays the status of the recharge. Depending on the color of the
LED the user can monitor the different status:
CHARGER LED COLOR
Yellow -
Yellow-
Orange -
Alternately green and yellow -
Green -
Alternately green and orange -
DOMU124_EN_v2_20221206_Poweo® 300 User manual
/
/
MODE
The battery is not connected to the charger
Reset and battery analysis
Fast recharge
Compensation charge
Slow charge/end of recharge
Error or battery overheat