Scaleo medical Poweo 300 User Manual page 41

Patient lifters
Table of Contents

Advertisement

Directives and manufacturer's declaration - electromagnetic immunity
The Poweo® is designed to be used in the electromagnetic environment specified below. The user should ensure it is used in such an environment.
Inmmunity test
Test level - IEC 60601
Electrostatic
± 8 kV contact
discharges
± 15 kV air
(E.S.D.)
IEC 61000-4-2
Transitional burst
± 2 kV for power supply
IEC 61000-4-4
lines
± 1 kV for input/output lines
Transitional surge
± 1 kV between phases
IEC 61000-4-5
± 2 kV line to ground
Voltage dips, short
<5% UT (> 95% dip in UT)
interruptions and
for 0.5 cycle 40% UT
voltage variations
(60% dip in UT) for 5 cycles
on the input lines
70% UT (30% dip in UT)
Power
for 25 cycles <5% UT ( >95%
IEC 61000-4-11
dip in UT) 5 s
Magnetic field @
30 A/m
50/60 Hz
IEC 61000-4-8
NOTE U
is the AC mains voltage prior to application of the test level.
T
Directives and manufacturer's declaration - electromagnetic immunity
The Poweo® is designed to be used in the electromagnetic environment specified below. The user of the Poweo® must ensure it is used in such an
environment.
Conducted
3 Vrmsf
RF
150 kHz to 80 MHz
disturbances
IEC 61000-4-
6
3V/m 80MHz to
2.7GHz with
Radiated RF
discreet
disturbances
frequencies 385,
IEC 61000-4-
450, 710, 745,
3
780, 810, 870,
930, 1720, 5240,
5500, 5785, 1845,
1970, 2450 with
level between
9V/m and
28V/m
NOTE 1: At 80 MHz and 800 MHz frequency range applicable is the highest.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
41
Level of compliance
± 8 kV contact
± 15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
Portable and mobile RF communications should not be used closer to any part of the Poweo®, including
cables, than the recommended separation distance calculated from the equation applicable to the
frequency of transmitter. Recommended separation distance
Non applicable
Non
applica
ble
d= 0,35√P 80MHz to 800MHz
d= 0,7√P de 800MHz to 2,5GHz
3 V/m
Where P is the maximum power output of the transmitter in watts (W), according to the transmitter
manufacturer and d is the separation distance in meters (m). The field strengths from fixed RF
transmitters, as determined by an electromagnetic investigation (a) should be below the level of
compliance in each frequency range. (b) Interference may occur in the vicinity of the equipment marked
with the following symbol:
DOMU124_EN_v2_20221206_Poweo® 300 User manual
Electromagnetic environment guidelines
Floor materials should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, should the relative humidity be
at least 30%.
Not applicable
Not applicable
Not applicable
Quality of power supply network should be of a typical commercial
or hospital environment. If the user of Poweo® requires continued
operation during network outages of power, it is recommended to
feed the Poweo® from an uninterruptible power supply or battery.