Understanding the information of the remote
The maintenance LED indicator is used to indicate a possible malfunction AND the periodic
inspection status. Therefore, it can display multiple consecutive status.
Example: in
LED displays alternately flashing RED and solid ORANGE.
Managing the maintenance LED status is easy:
Continuous solid light = default.
The light is continuous and the lifter is not operating = main fuse replacement is
necessary.
Flashing light = periodical inspection management.
According to the EN 10535 standards the time period between periodical inspection default values
is 12 months. The periodical inspection time setting can be changed using the Poweo® service tool
software.
BATTERY LED
– Solid green = more than 10 cycles are still
possible.
– Solid red: = less than 10 cycles left before
battery empty status.
– Flashing red: battery alarm (only one
cycle left before the battery is completely
empty).

Choosing the sling

The recommended slings for the Poweo® 300 are the slings listed in the "Accessories" section of
this manual, which are manufactured by SCALEO Medical. SCALEO Medical slings are Class I
medical devices (European Regulation 2017/745) and are compliant to the ISO EN 10535 European
standard.
19
case
of
default AND periodical inspection
service tool software or call us for maintenance.
performed within 2 months.
been performed in time.
DOMU124_EN_v2_20221206_Poweo® 300 User manual
MAINTENANCE LED
– Solid orange: default. Use the Poweo
– Flashing orange: periodic inspection must be
– Flashing red: periodic inspection has not
date
exceeded, the
®