Before first use, the battery must imperatively be 100% charged. Please refer to the "Battery
management" section for this operation. Your lifter will be ready for use as soon as the battery
is charged.
Your lifter will be usable as soon as the battery is charged
Whenever possible, keep the carton box and packing accessories for later
transportations.
Commissioning the Poweo
SCALEO Medical provides a universal power supply adapted to any voltage/frequency
value. Before charging the battery, you will need to check that the plug provided fits
to your local AC outlet.
If the place of use was informed before ordering, SCALEO Medical will provide the right
type of AC cord.
Before each use check the correct installation of the lifter and of its accessories, verify proper
operation before use with a patient.
The control points presented below only take a few minutes and allow you to ensure a safe
®
use of your Poweo
patient lifter:
Make sure that no part is damaged or show signs of wear,
Check the tightness of all the safety screws and pins,
Inspect accessible screws and make sure none are loose,
Check the lights on the remote control: battery indicator should be solid green;
maintenance indicator should be off,
Check the features: raise and down the arm, leg opening and emergency stop.
In case of malfunction or doubt about the proper operation, do not use the Poweo®
patient lifter. Inform your technical service, your distributor, or contact us at the
address on the back of this manual.