Options Configuration; Insite Configuration; Pnf Configuration; Pnf Gui Control Area - GE Healthcare Optima XR200amx Installation Manual

Hide thumbs Also See for Healthcare Optima XR200amx:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5336113-1EN, R
IRECTION

4.2.7 Options configuration

The Optima XR200amx/XR220amx systems are delivered with a base set of operating software
that is configured for immediate operation.
Optional software and hardware purchased by the site must be enabled by GE Healthcare service
personnel, such as:

4.2.8 InSite configuration

Contact a GE service representative to configure InSite ExC settings.

4.2.9 PNF configuration

The Firewall option at the top of the PNF area must be set to On. If it is set to Off, select TURN
FIREWALL ON.
After modifying any PNF options, select RESTART FIREWALL to apply the changes.

4.2.9.1 PNF GUI Control area

The PNF GUI Control area, near the top of the screen, contains these options:
Table 4-9 PNF GUI Control options
Filter Settings
Backup/Restore
Network Tools
Configure PNF
Page 64
10
EVISION
Wireless network connectivity
DAP (Dose Area Product) meter (standard with Optima XR220amx, optional for Optima
XR200amx)
Repeat Reject Analysis (RRA) (Optima XR220amx or Optima XR200amx with digital upgrade
only)
Auto Protocol Assist (APA) (Optima XR220amx or Optima XR200amx with digital upgrade
only)
Generator Power - 30kW
When the firewall is off, all network traffic in both directions is allowed. This setting is used
primarily for troubleshooting.
When the firewall is on, it allows only the network communications that are specified by its
configuration and rejects all the rest.
Option
Configure filter settings that allow access through the firewall by
adding allowed services and IP addresses, and removing filters.
Options:
BACKUP CURRENT FILTERS: Create a backup of the current filters.
RESTORE FROM BACKUP FILTERS: Restore the filters to a
previously backed up set.
RESTORE TO FACTORY DEFAULTS: Reset all filter settings to
those originally installed.
REMOVE BACKUP FILTERS:
Show the firewall rules that are currently in effect on this system.
Select REFRESH to refresh the list of rules.
Select which red interfaces the filter settings should be applied to.
Select UPDATE RED INTERFACES to apply the changes.
Section 4.2 Using the System Configuration Tool (SUIF)
O
XR200
PTIMA
AMX
I
NSTALLATION
Action
/XR220
AMX
M
ANUAL