3.6.8 Installing the detector grid

The digital detector grid is integrated with a holder that fits the detector exactly. The grid fits over
the detector handle and has raised edges to fit around the detector. See
Item
Figure 3-24 Detector grid

3.6.8.1 Unpacking and inspecting the grid

Inspect the surface of the grid (inside surface of the grid holder) for any signs of damage incurred
during shipment. Look for signs of scuffing, abrasions, or dents. The patient contact side of the grid
holder is covered with a padded overlay. Inspect it for any signs of scrapes or tears. Check the
perimeter of the grid holder for signs of sharp edges or damage/fraying to the carbon fiber structure.

3.6.8.2 Grid handling

Sensitive Device
The grid is manufactured with a delicate mechanical structure of thin lead strips, inter-spaced with
strips of aluminum. Do not subject it to drops, impacts or crushing conditions.
10
EVISION
To remove, use a small flat-blade screwdriver. Insert the screwdriver tip under the slot in the
side of the tag, and pry until the tag pops out.
Description
1
Grid (6:1 grid is a standard component, 8:1 grid is a purchasable option)
2
Detector
Chapter 3 Installation
O
XR200
/XR220
PTIMA
AMX
I
M
NSTALLATION
ANUAL
Figure
3-24.
Page 55