Recommended Materials And Equipment - GE Healthcare Optima XR200amx Installation Manual

Hide thumbs Also See for Healthcare Optima XR200amx:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5336113-1EN, R
IRECTION

1.3.5 Recommended materials and equipment

10
EVISION
Cover the work surface with approved static-dissipative material. Provide a wrist strap
connected to the work surface and properly grounded tools and equipment.
Use static-dissipative mats, foot straps, or air ionizers to give added protection.
Handle electrostatic sensitive components, parts, and assemblies by case or PCB laminate.
Handle components only at static-free work areas.
Turn off power and input signals before inserting and removing connectors or test equipment.
Use static-safe fixtures when fixtures must directly contact dissipative surfaces.
Keep work area free of non-conductive materials, like plastic assembly aids and Styrofoam.
ELECTRICAL SHOCK HAZARD
This section defines a workzone that is not connected to the system under service. The mobile
system is a battery powered unit that can deliver lethal current. Ensure you are not providing an
earth ground path when practicing ESD procedures.
Materials and equipment that are recommended for use in preventing static electricity include:
Anti-static tape, smocks, aprons, or sleeve protectors
Conductive bins, foam, tabletop workstations with ground cord of one-megohm +/-10%
resistance, and other assembly or soldering aids
Static-dissipative table or floor mats with hard tie to ground
ESD field service kits
Static awareness labels
Wrist straps and footwear straps providing one-megohm +/-10% resistance
Material handling packages:
-
Conductive plastic bags, plastic tubes and tote boxes
-
Opaque shielding bags
-
Transparent metalized shielding bags and shielding tubes
Chapter 1 Safety
O
XR200
/XR220
PTIMA
AMX
I
M
NSTALLATION
ANUAL