10. Mount the clamp to the Thorax chassis. Slide the clamp to the rear of the mounting screw slots.
11. Tighten the clamp around the cable bundle, ensuring that the cables are not pinched by the
Item
Figure 4-14 Correct cable bundle arrangement
12. Install the front bin, top cover and side covers.
13. Re-install the cable cover on the vertical column cable bundle. See the System manual,
14. Configure the system to enable the DAP meter option. See
15. Once the DAP meter option has been enabled, it must be 'switched on' at the Applications
Page 84
10
EVISION
clamp.
Description
1
DAP cable (optional for Brivo XR285amxOptima XR200amx systems)
2
Wireless cable (only valid for Optima XR220amx systems)
3
Vertical brake cable
4
Horizontal arm lock cable
Replacement Procedures chapter, Cable Cover Installation section.
page
64.
level:
a.
On the Applications desktop, click the UTILITIES button.
b.
On the Utilities desktop, select PREFERENCES > OPTIONS.
c.
Turn the DAP function ON.
d.
Refer to the system Operator Manual for instructions on DAP usage.
Section 4.10 Installing hardware options
O
C
4.2.7 Options configuration on
XR200
/XR220
PTIMA
AMX
AMX
I
M
NSTALLATION
ANUAL
2
1
3
4