Performing Loto - GE Healthcare Optima XR200amx Installation Manual

Hide thumbs Also See for Healthcare Optima XR200amx:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5336113-1EN, R
IRECTION

1.2.2 Performing LOTO

Page 24
Section 1.2 Lock Out/Tag Out (LOTO) procedure for electrical power
10
EVISION
1.
Notify all affected personnel working in the area that LOTO is being performed.
2.
Exit all system software.
3.
Turn off the Optima XR200amx/Optima XR220amx system by pressing the power button and
holding it for several seconds until the power button light ring turns blue, indicating that the
system is shutting down.
4.
When the power light ring begins to blink blue, indicating that the system is in standby, switch
the circuit breaker (E2) to the OFF position. See
Figure 1-3 Circuit breaker (E2) in OFF position
5.
Unplug the power cord from the wall outlet.
6.
Apply an approved AC power plug locking device to the AC power plug (E1), then apply your
personal red lock and tag. See
Figure 1-4 AC power plug with locking device applied
7.
Wait three minutes for discharge of stored energy.