Table 4-4 General tab parameters
Application Language
QAP Timeout (Optima
XR220amx or Optima
XR200amx with digital
upgrade)
VA Site
Screen Blank Timeout
Image Mask Adjustment
(Optima XR220amx or
Optima XR200amx with
digital upgrade)

4.2.4 Hardware configuration

Table 4-5 Hardware tab parameters
Collimator Light Maximum
On Time
Collimator Lamp Install
Date

4.2.5 Generator configuration

Table 4-6 Generator tab parameters
Tube HUR Alarm
Tube ID
Tube Serial
Tube Install Date and
Time
Page 60
10
EVISION
Parameter
Language used for application screens.
Interval between QAP tests, in days. When time-out is reached,
system users see a warning flag indicating that a QAP test should be
performed.
Editable by GE Healthcare personnel only.
Applies to USA Veterans Affairs facilities.
Yes or No.
Period of inactivity (no user input or processing activity) after which
the screen will go blank, in minutes.
Adjust the mask that is applied to the acquired image from the
detector. Options: 3mm, 2mm, 1mm, 0, -1mm, -2mm, -3mm.
Parameter
Time, in seconds, that the collimator light will remain on.
Date the collimator lamp was installed. YYYY-MM-DD format.
Parameter
Percentage of the tube thermal capacity limit, from 0 to 100. When this
limit is reached, X-ray exposure is inhibited and an alarm sounds.
Tube ID. Not editable.
Serial number of the tube. Confirm the serial number matches that of
the attached tube.
Date and time the tube was installed. YYYYMMDD HH:MM:SS
format.
Section 4.2 Using the System Configuration Tool (SUIF)
O
XR200
PTIMA
AMX
I
NSTALLATION
Definition
Definition
Definition
/XR220
AMX
M
ANUAL