7
Specifications and Characteristics
Specifications
All specifications are warranted. Specifications are valid after a 30-minute warm-up period
NOTE
and ±5 °C from last calibration temperature.
Table 5
Name
Bandwidth (-3 dB)
DC vertical gain accuracy:
Timebase accuracy:
Trigger sensitivity Ch 1, 2, 3, 4
(DC coupling):
1
138
Specifications
1
:
20 MHz when vertical scale is set to < 5 mV (1X probe attenuation).
Value
DSO1024A, DSO1022A: 200 MHz
DSO1014A, DSO1012A: 100 MHz
DSO1004A, DSO1002A: 60 MHz
2 mV/div to 5 mV/div: ±4.0% full scale
10 mV/div to 5 V/div: ±3.0% full scale
±50 ppm from 0 °C to 30 °C
±50 ppm + 2 ppm per °C from 30 °C to 45 °C
+ 5 ppm * (years since manufacture)
≥ 5 mV/div: 1 div from DC to 10 MHz, 1.5 div from 10 MHz to full
bandwidth
< 5 mV/div: 1 div from DC to 10 MHz, 1.5 div from 10 MHz to
20 MHz
Keysight 1000A Series Oscilloscopes User's Guide