Summary of Contents for Keysight Technologies 16048G
Page 1
Keysight 16048G/H Test Leads Operation and Service Manual...
Page 2
Declaration of Conformity the Software. With respect to any Keysight Technologies, Inc. as technical data as defined by FAR governed by United States and Declarations of Conformity for this 2.101, pursuant to FAR 12.211 and international copyright laws.
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the 16048G/H has been checked mechanically and electrically. The contents of the shipment should be...
Analyzer. Step 1. Set the 16048G/H to the test connectors on the front panel of the Impedance Analyzer by gradually coupling the four BNC connectors and fastening screws of the fixture with the test connectors and accessory mounting holes of the instrument until they come to complete contact.
2 Overview Product Overview The 16048G/H consists of a direct attachment, 4-terminal pair interface that is equipped with four BNC (f) connectors. These test leads are used to attach user-fabricated test fixtures. Cable length of the 16048G is 1 meter, the 16048H is 2 meter.
Overview Functions Functions Figure 2-2 shows name of each part of the 16048G/H and Table 2-1 shows their function. Figure 2-2 16048G/H Parts Table 2-1 16048G/H Function NAME FUNCTION Instrument Side Connect to the instrument’s UNKNOWN terminal. Connection Block Connection Block...
3. Press the Adapter [ ] key to bring up the Adapter Setup Menu. [ ] shows the current settings. 4. Select 4TP 1M in the case of 16048G, 4TP 2M in the 16048H. When the selection is completed, the softkey label will show Adapter [4TP 1m] or Adapter [4TP 2m].
Page 12
Phase [Done]. Press the [Save Phase] key. 8. Connect the Lcur, Lpot, Hcur, Hpot terminal on the 16048G or 16048H to the Lcur, Lpot, Hcur, Hpot terminal of the 100 Ω resistor respectively.
Page 13
Operation E4990A Setting 9. Press the E4990-61001 [-] key to start the load data measurement. When the load data measurement is completed, the softkey label changes to E4990-61001 [DONE]. 10.Press the Save Impedance key. Keysight 16048G/H Test Leads...
When you use the optional BNC Bracket (Keysight P/N 16048-60073) or an user fabricated test fixture, fixture compensation is required. Refer to the help manual of the Impedance Analyzer for the fixture compensation procedure. Keysight 16048G/H Test Leads...
Specifications Specifications 4 Specifications This chapter provides specifications of the 16048G/H Test Leads. Specifications Applicable Instruments E4990A Maximum Vol tage ± 42 V peak max. (AC+DC) Operating Temp. –20°C to +150°C Environment (Exclude up to 20 cm from four terminal pair...
Page 16
Specifications Specifications Keysight 16048G/H Test Leads...
“SG431xxxxx and below” Serial number for RoHS 16048G/H: “MY432xxxxx and above” / “SG432xxxxx and above” Maintenance An exploded view of the 16048G/H for parts identification is shown in Figure Figure 5-4. Due to limited availability of RoHS compliance station and technical difficulties in RoHS soldering, only parts and support level that do not require RoHS soldering are supported.
Page 18
Service Maintenance Figure 5-1 Replacable Parts (part 1 of 4) Keysight 16048G/H Test Leads...