Power meter range calibrator

Advertisement

Quick Links

Keysight 11683AZ Option
H01 Power Meter Range
Calibrator
Operating and
Service Manual

Advertisement

loading
Need help?

Need help?

Do you have a question about the 11683AZ Option H01 and is the answer not in the manual?

Questions and answers

Summary of Contents for Keysight Technologies 11683AZ Option H01

  • Page 1 Keysight 11683AZ Option H01 Power Meter Range Calibrator Operating and Service Manual...
  • Page 2 FAR 27.401 or DFAR 227.7103-5 (c), adhered to, could result in personal injury or as applicable in any technical data. death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. Keysight 11683AZ Option H01 Operating and Service Manual...
  • Page 3 ..........9 Keysight 11683AZ Option H01 Operating and Service Manual...
  • Page 4 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11683AZ Option H01 Operating and Service Manual...
  • Page 5: List Of Tables

    Digital voltmeter reading (OHM) ......10 Keysight 11683AZ Option H01 Operating and Service Manual...
  • Page 6 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11683AZ Option H01 Operating and Service Manual...
  • Page 7 Keysight 11683AZ Option H01 Power Meter Range Calibrator Operating and Service Manual Special Supplement Description Specifications Operation Performance Test...
  • Page 8: Specifications

    Special Supplement Description The 11683AZ Option H01 is a standard 11683A range calibrator that has been modified so that it will operate with either the Internal DC-Reference source, or an external programmable DC-Reference source. Note that the 11683AZ Option H01 is not equivalent to the standard 11683A. When ordering this NOTE product, you will need to order 11683AZ Option H01 instead of 11683A.
  • Page 9: Performance Test

    4 Connect the jumper between the DC REF INPUT BNC white wire and the RANGE switch green wire using an alligator test clip. 5 Set the REF SELECT switch to EXT. 6 Connect the 11683AZ-H01 Range Calibrator to the power line and turn it ON. Keysight 11683AZ Option H01 Operating and Service Manual...
  • Page 10 10 Disconnect the 4-wire cable and jumper from the 11683AZ-H01. This concludes the external DC-reference circuit performance test. If any of the voltage or resistance readings are incorrect, refer to the troubleshooting information in the 11683A Operating and Service Manual (11683-90014). Keysight 11683AZ Option H01 Operating and Service Manual...