Oem-Menu; Maximum Temperature - Thermo Scientific 50135178 AFT 3 Operating Instructions Manual

Barnstead pacific aft water purification system
Table of Contents

Advertisement

T

13.2 OEM-menu

Basic settings and limiting values can be changed in this menu.
To make changes in the OEM-menu, the system control must previously be unlocked (see
12.1.8)
Calling the OEM-menu:
Simultaneously pressing the INT-key and the Nonstop-Key calls the OEM-menu. Following
this, the prompt "OEM-menu Press enter!" appears. When this is confirmed with Enter, the
first menu point can be worked on. To simplify changing settings, press the UV-key to select
the position at which you want to change a number. Using the arrow keys now allows a
number from 0 to 9 to be entered at that position.
A press on the menu-key takes you to the next menu point.
The OEM-menu prompt display shows:

13.2.1 Maximum temperature:

A single press on the menu-key:
The maximum temperature the system can be exposed to can be set in this menu.
When this temperature is exceeded, the "Max. Temp." Fault message is triggered.
Settings above 50°C cause the limit evaluation to be suppressed, and the word "Off" appears
in the display. This is shown in the fourth line of the display.
Basic setting:
Setting range:
1 – 50 °C
The display shows:
30
OEM-Menu
Press enter
50 °C
OEM-Menu
max. Temp.
35 °C
Thermo Fisher Barnstead Pacific AFT
Pure water system

Advertisement

Table of Contents
loading