Oem-Menu; Maximum Temperature - Thermo Scientific LabTower AFT 20 Operating Instructions Manual

Water purification system
Table of Contents

Advertisement

14.2 OEM-menu

Basic settings and limiting values can be changed in this menu.
To make changes in the OEM menu , the system control must previously be unlocked (see
12.1.8).
Calling the OEM menu:
Simultaneous pressing of the Interval-key and the Nonstop-key calls the OEM menu. Following
this, the prompt "OEM menu Press enter" appears. When this is confirmed with the Enter-key,
the first menu prompt can be worked on. To simplify changing settings, press the UV-key to
select the number in a value which you want to change. Using the arrow keys now allows a
number from 0 to 9 to be entered at that position.
A press on the menu-key takes you ti the next menu prompt.
The OEM menu prompt display shows:

12.2.1 Maximum temperature:

A single press on the menu-key:
IThe maximum temperature to which the system can be exposed can be set in this menu. When
this temperature is exceeded, the "max. Temperature" fault message is triggered. Settings
above 50 °C cause the limiting value to be suppressed and the word „off" appears in the dis-
play. It is shown in the 4
Basic setting:
Setting range:
The display shows:
OEM menu
Press Enter !
th
line of the display.
50 °C
1 - 50 °C
OEM menu
max. Temperature
35 °C
Thermo Scientific Barnstead LabTower AFT
Pure water system
31