Operation
When a channel enabled for auto calibration is in the calibration
mode, a message indicating ZERO cal. or SPAN cal. will appear on the
measurement screen. For example, if channels 1 and 2 have the auto
calibration setting enabled, when the calibration event takes place, both
channels 1 and 2 will indicate ZERO cal. when the zero calibration is
being performed. Zero calibrations are performed first and all enabled
channels will zero simultaneously.
When the zero calibrations are finished, channel 1 will enter the
span calibration followed by channel 2.
3.5.4.4 R
S
EMOTE
TART
An auto calibration is always available by keeping the remote start
input closed for at least 1.5 seconds. The remote start can initiate an auto
calibration whether the parameter setting is ON or OFF.
3.5.4.5 F
R
ORCED
UN
Auto calibration can be performed manually or forcibly stopped
while the calibration is performed.
Teledyne Analytical Instruments
/S
A
C
TOP OF
UTO
ALIBRATION
Model 7600
62